Process Control Home
 
Alpha Controls
 

SmartReader Plus 4 Data Logger

SmartReader Plus 4
SmartReader Plus 4 Data Logger provides easy monitoring and troubleshooting of pneumatic control systems. The unit is completely self-contained with a 1/4" quick-release pressure fitting to allow a fast and trouble-free connection. As an added bonus: SmartReader Plus 4 can also monitor temperature internally, or temperature and RH externally. (RH sensors/probes and thermistor temperature probes sold separately).


Four Channel Pressure, Temperature and RH Data Logger
SmartReader Plus 4 

Calibrated Accuracy:

+/-0.5% of full scale @ 25ºC (77ºF)

Memory Size:

32 KB and 128 KB

No. of Channels:

Four. (One for ambient temperature, one for pressure, and 2 for the optional Relative Humidity/Temperature Probe)

Remote Network Communications:

See Network & Communications

Industry:

Pressure, Compressed Air, Airflow Monitoring, Air Quality

PDF

SmartReader,Plus,4,Data,Logger

Size: 107 x 74 x 22 mm (4.2" x 2.9" x 0.9")
Weight: 110 grams (3.75 ounces)
Case Material: Noryl® Plastic
Operating Limits: -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy: +/-2 seconds per day
Battery: 3.6 volt Lithium, 1 Amp-Hour
Power Consumption: 5 to 10 microamps (continuous)
Battery Life: 10-years warrantied (under normal use). Factory replaceable
Memory Size: 32 K (21,500 readings). 128 K (87,000 readings). 1.5 MB (1,000,000 readings). See specific models for availability
Sampling Method: 1. Continuous (First-in, First-out); not available with sample rates faster than 8 seconds.) 2. Stop when full (Fill-then-stop)
Sampling Rate: User selectable rates from 25 per second to one reading every 8 hours. (BP-101 battery pack or PS-201 Power Supply required for sample rates faster than 8 seconds. See Accessories section for details)
Resolution: 12 bit (1 part in 4096)
PC Requirements: IBM PC or 100% compatible running MS® Windows 3.1,95,98, or NT, with a least 2 MB RAM, 2 MB of hard drive disk space and one free serial port
Product Approvals: Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets Class A radiated and conducted emission requirements of Section 2 of the Radio Interference Regulations of Communications Canada. IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/ Burst and Radiated Emissions respectively for the Commission of the European Communities (CE)
Resistance to X-Rays: Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport X-Ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 - 28 curies @ 30 second exposure source to object distance 5"


Powerful Solutions for Instrumentation & Control Problems!
SmartReader Plus 4 Data Logger

Site Designed By Panther Marketing